Description
odel Number
Dimensions, D x W x H
Weight
Voltage requirements
Frequency requirements
Power requirements
Logic power backup outage
Equipment class
Degree of protection against
electrical shock
Enclosure protection against ingress
of liquid
Foot switch protection against
ingress of liquid
Mode of (electrical) operation
Transport temperature
Transport relative humidity
Transport atmospheric pressure
Storage temperature
Storage relative humidity
Storage atmospheric pressure
Operation temperature
Operation Humidity
Operation atmospheric pressure
Pollution
Installation/Overvoltage
Fuse
Cables
The AngioJet Ultra Console is UL LLC Classified to
ANSI/AAMI ES60601-1:A1:2012, C1:2009/(R)2012 and A2:2010/(R)2012,
CSA CAN/CSA-C22.2 No. 60601-1:14; UL 60601-1, CAN/CSA C22.2 No. 601.1.
Note: If it becomes necessary to replace the fuse, replace fuse with the type and
rating specified. Failure to do so may result in device damage or risk of fire.
5000A
25 in x 16.5 in x 54 in
(63.5 cm x 42 cm x 137 cm)
140 lbs (63.5 kg)
100/120/220/240 VAC
50/60 Hz
900 VA
60 seconds for conditions of power loss
Class 1
Defibrillation-Proof Type CF Applied Part
IPX 1
IPX 8
Continuous
-25 °C to 55 °C (-13 °F to 131 °F)
10% to 95% (noncondensing)
500 hPa to 1060 hPa
-25 °C to 55 °C (-13 °F to 131 °F)
10% to 95% (noncondensing)
500 hPa to 1060 hPa
10 °C to 40 °C (50 °F to 104 °F)
30% to 75% (noncondensing)
700 hPa to 1060 hPa
No greater than degre