- Height: 5.75 in (14.6 cm)
- Width: 16.75 in (42.5 cm)
- Depth: 10 in (25.4 cm),/li>
- Weight: 8 lbs (3.6kg),//li>
Environmental Operating Conditions
- Ambient Temperature: 50°F to 104°F (10°C to 40°C)
- relative Humidity: 10% to 75%, non-condensing
Environmental Storage Conditions
- Ambient Temperature: 14°F to 131°F (–10°C to 55°C)
- Relative Humidity: 10% to 90%, non-condensing
Power Requirements
- Nominal line voltage: 100-230 VAC
- Line frequency: 50/60 Hz (operates over 47-63 Hz)
- Power consumption: ≤30 VA
- Monitor DC input: 12 Vdc at 2.5A
Operating Specifications
- Technique Pulsed Doppler with auto correlation processing
- Transducer Type 9-crystal
- Pulse Repetition 2 kHz (all modes)
- Frequency Pulse Duration 92 μs
- Transmitter Frequency 1.151 MHz
- Spatial-Average Temporal Isata <5 mW/cm2
- Average Intensity
- Focal 20 dB Beam Area 16.6 cm2, at a range = 7 cm
- Peak Instantaneous 1.8 mW/cm2
- Intensity
- Heart Rate Counting Range 50-210 BPM
- Leakage Current Complies with IEC 60 1.1 and/or IEC 601.1.1 harmonized national standard; Complies with EN 60601-1-2:2002
FECG Mode
- Technique: Peak detecting, beat-to-beat cardiotachometer
- Heart range counting range: 30-240 BPM
- Heart rate resolution: 1 BPM
- Artifact elimination: service selectable, ±25 BPM artifact rejection
- Countable input signal range: 15µV to 2mV peak-to-peak
- Offset voltage tolerance (differential): ±300 mVdc maximum
- Maximum common mode voltage: 20V peak-to-peak
- Common mode rejection: Balanced: >120 dB at mains frequency, with patient cable, Unbalanced 5kΩRA or LA: >110 dB at mains frequency
- Input impedance: Differential: >10MΩ; Common mode: >20MΩ
- Mains frequency rejection: >40 dB
- Leakage current: complies with IEC 601.1 and/or IEC 601.1.1 harmonized national standard
- Isolation, mains-to-patient: >5656 Vdc
Ultrasound Mode
- Technique: pulsed Doppler with autocorrelation processing
- Transfer transducer type: 9-crystal
- Pulse repetition frequency: 2 kHz (all modes)
- Pulse duration: 92 µs
- Transmitter frequency: 1.151 MHz
- Spatial-Average temporal average intensity: Isata<5 mW/ cm²
- Focal 20 dB beam area: 16.6 cm², at a range = 7 cm
- Peak instantaneous intensity: 1.8 mW/cm²
- Heart rate counting range: 50-210 BPM
- Leakage current: complies with IEC 601.1 and/or IEC 601.1.1 harmonized national standard