GE E-miniC-00 C.O.2 Module With Mini D-Fend Water Trap

$695.00

In good physical condition, It has not been fully tested, We don’t know full functionality of the unit and unit is not guaranteed. Being sold AS IS no returns.

1 in stock

Categories: ,
Report Abuse

Description

Single-width Airway Module, E-miniC

For compact CO2 sidestream respiratory monitoring 

The E-miniC Module is a single-width, plug-in module. The module is intended for use for patients weighing over 5 kg (11 lb) only

Features

• CO2 side-stream measurement done at the patient’s airway
• Very small and reliable measurement technology

• Inspired and end-tidal CO2 values displayed on the screen as waveforms and in numerical format 

• Full numerical and graphical trends of all measured data

•Measurement of respiration rate up to 80 breaths per minute 

•Fast response time 

•Wide measurement range for CO2 

•Adjustable high and low alarm limits for EtCO2 and FiCO2 

•Manual adjustment for N2O and O2 allowed for improved CO2 measurement accuracy 

•Accommodates water separation system which is lightweight, and has good tolerance for tilting

Technical specifications

General

Sampling rate 150 ±25 ml/min 

Displayed values updated breath by breath 

Manual O2 and N2O compensation allowed to improve CO2 measurement accuracy • When manual O2 compensation is not activated (O2 concentration 40% to 95%), CO2 reading decreases <0.3 vol% 

•  When the manual compensation for O2 is activated (O2 concentration 40 to 95%), the error in the CO2 reading is < 0.15 vol% 

• When manual N2O compensation is not activated (N2O concentration 40 to 80%), CO2 reading increases < 0.8 vol% 

• When the manual compensation for N2O is activated (N2O concentration 40 to 80%), the error in the CO2 reading is < 0.3 vol% 

Non-disturbing gases 

Ethanol, acetone, methane, nitrogen, water vapor: Effect on readings at 5.0 vol%: < 0.2 vol% (max)

Carbon dioxide (CO2)

EtCO2 End-tidal CO2 concentration 

FiCO2 Inspired CO2 concentration 

Measurement range 0 to 20 vol% 

(0 to 20 kPa, 0 to 150 mmHg) 

Accuracy CO2 concentration 0 to 15 vol% 

±(0.2 vol% + 2 % of reading) 

CO2 concentration 15 to 20 vol% 

±(0.7 vol% + 2 % of reading) 

Rise time < 300 ms with nominal flow 

Adjustable low and high limits for EtCO2 or FiCO2

Respiration rate (RR)

Measurement range 4 to 80 breaths/min 

Breath detection 1% variation in CO2 level 

Accuracy ±1/min in the range 

4 to 20 breath/min 

±5% in the range 

20 to 80 breath/min 

Resolution 1/min 

Adjustable low and high limits for respiration rate; alarm for apnea

Monitor compatibility

CARESCAPE modular monitors 

AS/3, CS/3 and S/5 modular monitors using software L-ANE02(A), L-ICU02(A) or later versions

Performance specifications

Functional alarms Blocked sample line, water trap replacement, water trap check, low gas sample flow 

Environmental specifications

Operating conditions 

Temperature 10 to 40°C (50 to 104°F) 

Relative humidity 10 to 95% non-condensing 

Atmospheric pressure 

variation 500 to 800 mmHg

  (666 to 1060 mbar)

Storage conditions

Temperature -25 to 70°C (-13 to 158°F) 

Relative humidity 10 to 95% non-condensing

Physical specifications

Dimensions (H x W x D) 11.2 x 3.7 x 20.9 cm 

 (4.4 x 1.5 x 8.2 in) 

Module weight 0.4 kg (0.9 lb)

Ready to ship in 1-3 business days from United States (US)


 

Shipping Policy

All of our products are professionally packaged to assure your items arrive safely.

Product Enquiry

Your personal data will be used to support your experience throughout this website, to manage access to your account, and for other purposes described in our privacy policy