Description
Display
Type: High resolution color TFT display.
Resolution: 800 X 600 pixels.
Size: 12.1″ configurable screen for max of eight waveforms.
Trace Speed: 12.5, 20 OR 50 MM/SEC for ECG, SpO2 & IBP 6.25, 12.5 OR 25 MM/SEC for respiration & CO2.
Optical Encoder Keys
Optical Encoder: Rapid access to all the functions and setting through single knob.
Hot Keys: Quick action hot keys for acknowledge, NIBP start/stop, record start/stop, IBP zero, freeze, stand by, go-to, CO2 pump off and home key.
3/5 Lead ECG Monitoring
Automatic detection of 3 or 5 lead ecg cable. Display of I, II, III with 3 – lead cable and I, II, III AVR, AVL, AVF andV leads with 5 lead cable. Built-in cautery and defibrillator protection. Optional st and arrhythmia analysis.
HR Range: 20 to 350 BPM
Respiration
Principle: Impedance pneumography. Display of respiration waveform with respiration rate.
RR Range: 4 to 150 BPM
Simultaneous monitoring of two temperatures. Display of T1 ANand T2 or ΔT (if difference between T1 & T2 is less than 9.9°C)
Probe: Compatible with YSI 400 series temp. probe(optional)
Temp. Range: 0°C to43°C
Pulse Oximetry (Nellcor module)
Pitch Variation: With change in oxygen saturation SpO2 range: 0-100%
Display: Plethysmograph, pulse strength & SpO2 values.
Accuracy: Adult- 100 TO 70% (+/- 2 digits), 0 to 69% (unspecified) (+/- STD. DEV.)
Neonate: 100 TO 70% (+/- 3 digits), 0 TO 69% (unspecified)
PR Range: 20 TO 230 BPM +/- 3 BPM (+/- STD. DEV.)
Invasive Blood pressure
Simultaneous monitoring of max. 4 channels. Display of systolic, diastolic and mean pressure. Individual and common channel zeroing. Labeling facility with automatic limits/gain/color setting. Pressure channel overlapping option available.
Range: -40 to 300 MMHG
IBP Scale: 5, 10, 15, 20, 30, 60, 80, 100, 120, 150, 180, 200, 240, 300 & auto
Non-Invasive Blood Pressure (Suntech Module)
Principle: Oscillometric
Display: Systolic, diastolic & mean pressure
Modes: Manual, stat (5 MIN), auto. (2 MIN TO 90 MIN)
Range: 20 TO 250 MMHG
Accuracy: +/- 3 MMHG
Capnography (Oridion Module)
Microstream capnography with display of CO2 waveform and digital values of EtCO2, FiCO2, & RR, CO2 ets priority for calculating resp. rate
EtCO2 Range: 0 – 99 MMHG
FiCO2 Range: 0 – 20 MMHG
Gain: 20, 40, 60, 80 MMHG
Flow Rate: 50 ML/MIN (-7.51 / + 15ML/MIN)
Warm-Up Time: 40 SECS (typically)
Units: MMHG, KPA / VOL%, RR
Range: 0 -150 BPM
Alarm settings for all parameters
Patient Alarm: Red flashing with audio and message indicator alarm tuo-set facility provided
Equipment Alarm: Yellow flashing with audio and message indicator
Alarm Suspend: Continuous yellow
Re-Alarm Facility: 3 min. for acknowledge alarm
Trends
72 HRS/100 Patients graphical and tabular trend withh zoom facility
NIBP Trend: 240 Reading tabular
Alarm Recall: Last 24 alarm events with date, time & message
Recorder (optional)
Type: Built-in channel thermal recorder for recording two waveforms (selectable) with numerical data and alarm status. nibp and tabular trend recording with patient information
Modes: Direct/delayed/continous ans record on alarm facility(selectable)
Dimension: (H X W X D) 208 X 320 X 158 MM
Weight: 6KG (with built-in battery & thermal array recorder)
Power Supply
Universal Power: 95 – 265V AC, 50/60 Hz
Current Rating: 0.7 A (Max)
Battery Type: 14.8V (4 AH/8 AH) Lithium Ion Batteries
Operating Time: Built-in rechargeable battery for 2/4 hrs of continuous operation under full charge condition (typically)
Charging Time: 8 HRS (typically)