Nonin Life Sense Capnography Pulse Oximeter Monitor Model LS1-9R

$275.00

In good physical condition, Powers on but has not been fully tested, full functionality is unknown and not guaranteed. Being sold AS IS no returns. We have multiple in stock the one you see in the photos may not be the one you receive.

1 in stock

Report Abuse

Description

Introduction 

About LifeSense

LifeSense allows healthcare professionals to non-invasively monitor pulse oximetry and capnometry on either intubated or spontaneously breathing patients. This very useful combination serves as a reliable indication of the patient’s respiratory and ventilation status. 

When measuring EtCO2, the patient is attached to the monitor by a sample line that can be an airway adapter for an endotracheal tube, a nasal cannula, or a nasal cannula with supplemental oxygen delivery. A variety of sample lines can be used and connected to a specially designed moisture trap, which is easily snapped into the slot on the left side of the monitor. The sampling lines or cannulas can be used with or without Nafion® tubing. Pulse rate and SpO2 are measured by a Nonin-branded PureLight® finger clip sensor, provided with the system. Use only those accessories and replacement parts recommended by Nonin. Refer to the “Accessories” section for more information.

LifeSense has visual and audible alarms when limit readings are outside the predefined limits. Limits can easily be adjusted using the touch panel display. The operator can pause or resume the alarm by pressing the Audible Alarm Pause/Resume button. 

LifeSense has a touch panel display where settings and adjustments are made. The touch panel display also shows battery status and fault messages. The only buttons on the monitor, ON/OFF and Audible Alarm Pause/Resume, are located on the upper right corner of the front panel. Next to these buttons there is a small indicator that turns green when the monitor is connected to a power outlet. LifeSense operates on battery power for approximately 8 hours.

Indications for Use

The LifeSense Model LS1-9R Capnography/Pulse Oximeter monitor is indicated for use in simultaneously measuring and displaying functional oxygen saturation of arterial hemoglobin (SpO2), end tidal carbon dioxide (EtCO2), respiration, and pulse rate of well or poorly perfused adult, pediatric, and infant patients. It is intended for use in environments where patients require continuous, non-invasive monitoring of these parameters by a healthcare professional (e.g., hospitals, medical facilities, post-operative care, within facility patient transport or home use. 

Warnings  

Do not use LifeSense in an MR environment or in the presence of flammable anesthetics or gases.

Do not use LifeSense during defibrillation.

LifeSense is intended only as an adjunct in patient assessment. It must be used in conjunction with other methods of assessing clinical signs and symptoms.

LifeSense is not intended to be used as a primary diagnostic apnea monitor.

Verify all alarm settings and limits during system start up to ensure that they are set as intended.

A hazard can exist if different presets are used on multiple LifeSense monitors in one care area.

To prevent improper performance and/or patient injury, verify compatibility of the monitor, sensor(s), and accessories before use.

No modifications to the device are allowed as it may affect device performance.

Inspect the pulse oximeter sensor application site at least every 6 to 8 hours to ensure correct sensor alignment and skin integrity. Patient sensitivity to sensors and/or adhesive strips may vary due to medical status or skin condition.

Avoid excessive pressure to the sensor application site as this may cause damage to the skin beneath the sensor

Never allow liquids to enter into or to be spilled onto the monitor. If liquid has penetrated into the monitor it must be checked by Nonin Technical Service.

Accessories marked “single-use” must be used on one patient only and be disposed of after usage.

To avoid patient injury, use only Nonin-branded PureLight® pulse oximeter sensors. These sensors are manufactured to meet the accuracy specifications of Nonin pulse oximeters. Using other manufacturers’ sensors can result in improper pulse oximeter performance.

Only use power supplies that either are supplied with LifeSense or specified by Nonin (see “Accessories”).

When selecting a sensor application site use an extremity without a catheter, blood pressure cuff, or intravascular infusion line.

Do not use a damaged sensor. If the sensor is damaged in any way, discontinue use immediately and replace the sensor. 

Misuse or improper handling of the pulse oximeter sensor could damage the sensor or the cable, which may lead to inaccurate readings. Never alter or modify the sensor since this may affect the performance or accuracy.

If the LifeSense fails to respond as described, discontinue use and contact Nonin Technical Service.

Use only Nonin recommended accessories and replacement parts (see “Accessories”).

LifeSense displays a BATT LOW message when it has approximately 20 minutes of use remaining before it shuts itself off.

EtCO2 value will be diluted when used in combination with supplemental oxygen. To get a true EtCO2 reading it is recommended that the supplemental oxygen is disconnected for a few seconds.

Oximeter readings may be affected by the use of an electrosurgical unit (ESU).

This device should not be used adjacent to or stacked with other equipment. If adjacent or stacked use is necessary, the device should be observed carefully to verify normal operation.

Prior to connecting LifeSense to the power supply and the power outlet, be sure to verify the voltage and frequency rating on the power supply are the same as the power outlet. If this is not the case, do not connect the monitor and power supply to the power outlet.

Ensure that all alarm volumes are audible in all situations. Do not cover or obstruct any speaker openings.

The use of accessories other than those specified in this manual may result in increased electromagnetic emission and/or decreased immunity of this device.

As with all medical equipment, carefully route patient cables and connections to reduce the possibility of entanglement, strangulation, or injury to the patient.

When turning on the monitor, verify that a beep is heard each time a button is pressed. If a beep is not heard, do not use the device. The speaker may not be functioning properly.

Cautions  

LifeSense should only be operated by trained licensed practitioners

To prevent damage to the monitor, operate and store the monitor in an upright position.

Setting alarm limits to extremes can render the alarm system useless. 

Secure LifeSense with mounting hardware if used in transport vehicles.

Do not mount LifeSense directly above the patient. If the monitor is mounted, be sure to check that the adjustable mounting clamp is securely affixed.

When mounting the monitor to a mobile pole, mounting the monitor higher than 1.5 meters (5 feet) or mounting more than 2 kilograms (4.5 pounds) of equipment onto the pole may result in tipping, damage to the equipment, or injury.

Always turn off the monitor prior to cleaning the monitor or changing the pulse oximeter sensor or moisture trap and/ or filter.

Portable and mobile RF communications equipment can affect medical electrical equipment.

The sample line, moisture trap, and filter are single-use disposable components. Do not disassemble the plastic parts of the single-use disposable moisture trap. Dispose all components in accordance with your local, state or national regulations regarding waste management.

Ear Clip and Reflectance SpO2 sensors are not recommended for pediatric or infant/neonatal use. The accuracy of these sensors has not been established for pediatric or infant/neonatal use.

This equipment complies with IEC 60601-1-2 for electromagnetic compatibility for medical electrical equipment and/ or systems. This standard is designed to provide reasonable protection against harmful interference in a typical medical installation. However, because of the proliferation of radio-frequency transmitting equipment and other sources of electrical noise in healthcare and other environments, it is possible that high levels of such interference due to close proximity or strength of a source might disrupt the performance of this device. Medical electrical equipment needs special precautions regarding EMC, and all equipment must be installed and put into service according to the EMC information specified in this manual.

Set or adjust alarm parameters one at a time.

Do not cover or block speaker opening. This may significantly reduce the sound volume.

Before each use, it is the operator’s responsibility to verify that the alarm limits are appropriate for the patient being monitored.

Always clean the Nonin PureLight reusable sensor after each patient use. Before cleaning, unplug it from the monitor.

The patient’s nasal passage may dry out if continuous monitoring is required. Check patient on a regular basis for nasal comfort.

If the EtCO2 value is out of normal range (4.4 – 5.7 Vol%/KPa or 33 – 43 mmHg) an internal air leak is possible. Replace the single-use disposable moisture trap and repeat the calibration procedure. If the problem persists, contact Nonin Technical Service.

If LifeSense is intended to be stored for longer periods of time, always charge the battery to full capacity before storing it in order to prevent damage to the equipment.

Avoid rapid temperature change or extreme temperatures. This can cause malfunction.

Never store or transport LifeSense where condensation can occur. However, if this has occurred, wait until all condensation has evaporated before using LifeSense.

Do not attempt to replace the battery inside the monitor. The battery is not field replaceable and cannot be replaced by the operator. Use only Nonin-specified components. Contact Nonin Technical Service when the battery needs replacing. Battery replacement by inadequately trained personnel could result in a hazardous situation.

Do not charge Li-Ion batteries at a temperature of 0 °C (32 °F) or less as this may result in significantly reduced battery life.

The sample line, moisture trap, Nafion tubing, filter, and T-connector are single-patient use, disposable components. Dispose all components in accordance with your local, state, or national regulations regarding waste management.

Do not sterilize or autoclave the monitor or sensors. Do not immerse in liquids. Do not disassemble the plastic parts of the single-use disposable moisture trap.

Never open the monitor housing/case. By opening the case you render your warranty invalid.

The oximeter sensor might not work on cold extremities due to reduced circulation. Warm or rub the finger to increase circulation, or reposition the sensor

In compliance with the European Directive on Waste Electrical and Electronic Equipment (WEEE) 2002/96/EC, do not dispose of this product as unsorted municipal waste. This device contains WEEE materials; please contact your distributor regarding take-back or recycling of the device. If you are unsure how to reach your distributor, please call Nonin for your distributor’s contact information.

This device is designed to determine the percentage of arterial oxygen saturation of functional hemoglobin. Factors that may degrade pulse oximeter performance or affect the accuracy of the measurement include the following:

-excessive ambient light                                                        – inadequate signal

– excessive motion                                                                – venous pulsations

– electrosurgical interference                                                – anemia or low hemoglobin concentrations 

– blood flow restrictors (arterial catheters, blood pressure    – cardiogreen and other intravascular dyes

  cuffs, infusion lines, etc.)                                                   – carboxyhemoglobin

– moisture in the sensor                                                       – methemoglobin

– improperly applied sensor                                                 – dysfunctional hemoglobin

– incorrect sensor type                                                         – artificial nails or fingernail polish

Each time the system is turned on, all audible alarms are disabled for 2 minutes unless the operator presses the Audible Alarm Pause/Resume button.

Capnography alarms are not active until the first breath is detected. 

Oximetry alarms are not active until the first pulse is detected. 

A functional tester cannot be used to assess the accuracy of a pulse oximeter monitor or sensor.

Be careful not to drop LifeSense on the floor or strike it against hard surfaces. If such an incident happens, do not use LifeSense until a functional test has been carried out.

All parts and accessories connected to the serial port of this device must be certified according to at least IEC Standard EN 60950 or UL 1950 for data-processing equipment.

The monitor is equipped with automatic barometric pressure compensation. End tidal pCO2 values displayed are calculated based on an atmospheric pressure of 760 mmHg and pH2O of 47 mmHg (example: 760 – 47 = 713, 713 x 5% = 36 mmHg).

Water or other liquid in the sampling tube may cause erroneous CO2 readings.

Ensure that all connections are tight, leak-free, and properly attached.

If the Nafion tubing becomes contaminated or damaged during use, discard it and replace it with a new one.

Radios and cell phones or similar devices may affect the LifeSense and should be kept at least 2.5 meters (8 feet) away from the device. Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast towers and TV broadcast towers may affect accuracy.

Failure of a network data coupling (serial cable/connectors) will result in loss of data transfer. 

Before first use and before storing this monitor, fully charge battery

If not in continuous use, fully charge battery at six-month intervals

Before first use and before storing this monitor, fully charge battery.

If not in continuous use, fully charge battery at six-month intervals.

 

Ready to ship in 1-3 business days from United States (US)


 

Shipping Policy

All of our products are professionally packaged to assure your items arrive safely.

Product Enquiry

Your personal data will be used to support your experience throughout this website, to manage access to your account, and for other purposes described in our privacy policy